The Internet Journal of Minimally Invasive Spinal Technology™ ISSN: 1937-8254

| Home | Editors | Current Issue | Archives | Instructions for Authors | Disclaimer | Bookmark and Share Share with others |
 

The Internet Journal of Minimally Invasive Spinal Technology is the official online journal of ISMISS/SICOT (International Society of Minimally Invasive Spinal Surgery, affiliate of SICOT) and AAMISMS (American Academy of Minimally Invasive Spine Surgery and Medicine)

Ijmist1Ijmist2Ijmist3Ijmist4

Miniature Robotic Guidance In MIS Spinal Surgery - Accuracy, Radiation And Procedure Time

Printer Read printer friendly
 Subscribe in a reader
Bookmark and Share Share with others
 

M. Hardenbrook MD
The Boston Spine Group Physical Address

A.J. Khanna MD
Johns Hopkins University Physical Address

I. Lieberman MD
Cleveland Clinic Foundation Physical Address

J. Wang MD
University of California, Los Angeles Physical Address

R.D. Guyer MD
Texas Back Institute Physical Address

Citation: M. Hardenbrook, A. Khanna, I. Lieberman, J. Wang & R. Guyer : Miniature Robotic Guidance In MIS Spinal Surgery - Accuracy, Radiation And Procedure Time . The Internet Journal of Minimally Invasive Spinal Technology. 2008 Supplement I - to IJMIST Vol 1 No 2


 

Abstract

Design: Prospective, controlled, multi-center, cadaveric study, evaluating efficacy of miniature robotic guidance for percutaneous pedicle screws.
Methods: Twelve cadavers, 234 screws, 17 surgeons (T9-S1). Accuracy (post-procedural CT), fluoroscopy time, radiation exposure (badge and ring dosimeters) and procedure time were measured.
Results: Mean accuracy was 2.6 mm (std 0.7mm) for control; 1.1 mm (0.4) for study group. Control averaged 136 mrem X-Ray exposure per surgeon; 4.2 in study group. 13 of 15 surgeons in study group (87%) experienced radiation levels below measurable threshold. Average radiation dose for control was 10.1 mrem/screw, with average fluoro time/screw 27.6 sec; 0.2 mrem and 1.2 sec respectively in study group. Average procedure time for control was 1:59 hours (ave. 19 screws), compared with 1:14 hr (20 screws) for new robotic users and 0:49 hr for experienced (18 screws). Average time per screw was 6:16 minutes (std 3:03 min), 4:03 minutes (1:05 min) and 2:45 minutes respectively.
Conclusions: Use of miniature, bone-mounted robotic guidance for percutaneous pedicle screws resulted in accuracy twice better than free hand. 56 times lower radiation and 23 times shorter fluoroscopy were observed, with time savings of 36% - 56%. Across all measures, robotic guidance improved performance and reduced variations between surgeons.