The Internet Journal of Orthopedic Surgery™ ISSN: 1531-2968| Home | Editors | Current Issue | Archives | Instructions for Authors | Disclaimer |The Utilization Of Trabecular Metal In Combination With Autgraft Bone In Acetabular Reconstruction: A Preliminary Sheep Model
Arash Nabavi
John R. Field
Citation: A. Nabavi & J. Field : The Utilization Of Trabecular Metal In Combination With Autgraft Bone In Acetabular Reconstruction: A Preliminary Sheep Model. The Internet Journal of Orthopedic Surgery. 2009 Volume 13 Number 2 Table of ContentsAbstractBackgroundWhen confronted with acetabular bone loss at the time of primary or revision hip replacement, there are three surgical options available; the utilization of a larger cup, the use of a cage in combination with a cemented cup or the use of impaction grafting techniques. Impaction grafting is not without problems however its primary advantage is a restoration of bone stock. The current pilot study has sought to evaluate the efficacy of trabecular metal used in combination with impaction grafting of autograft bone in the repair of a simulated Type III acetabular defect in a sheep model.MethodsA unilateral hip replacement was performed using modular cemented components. An acetabular defect was created in the cranio-dorsal wall of the acetabulum and impacted with harvested and milled autograft bone. A trabecular metal plug was applied to the graft bed prior to the cementing of a polyethylene cup using 1st generation cementing techniques. End-points measured included radiography, macrophotograhy, micro-radiography and histology.ResultsThe planned three month duration of the in vivo assessment was interrupted at 8 weeks due to recurrent dislocation. At that time, appropriate tissues were retrieved for assessment.Macrophotographs revealed little or no fibrous tissue along the interface between the trabecular metal and underlying graft bed. Microradiographs confirmed these findings. Fluorescent microscopy and light microscopy highlighted significant amounts of new bone formation within the trabecular metal impant along its interface with the underlying autograft bone.ConclusionsActive new bone formation was observed penetrating as much as 1-1.5 mm into the trabecular metal implant at 8 weeks following implantation. Over a longer period of time, with continued ingrowth and increasing strength, the provision of a stable mechanical environment is anticipated in response to the surgical techniques employed in this model. This preliminary study has demonstrated that trabecular metal can be successfully used in impaction grafting when it is not in contact with peri-prosthetic host bone. IntroductionPrimary or secondary acetabular bone loss, at the time of initial total hip replacement (developmental dysplasia), or revision of actebalular components with cavitatory defects in situ, continue to be problematic. Currently, there are three preferred techniques used when dealing with Type III (1-4) acetabular defects; these include over-reaming and the utilization of a larger cup (5,6), the use of a cage in combination with a cemented cup (7,8) or impaction grafting techniques (9-15). The use of large revision cups, although technically easier, has some major disadvantages including a net loss of bone stock which is further complicated by stress shielding exacerbated by the larger, stiffer cups used (5,6). The utilization of retainment cages in combination with cement has been reported to have a 20% failure rate at 5 years (16,17). Likewise, impaction grafting has problems including cup ‘spin-out’ because of a failure to integrate at the bone graft-cement interface (14,16). Concurrent problems include inadequate initial stability and difficulty controlling cement intrusion into the bone graft. However, the advantages include augmentation of bone stock and a restoration of the centre of rotation of the hip (17-19). Slooff et al (1984) first reported the successful utilization of impaction grafted bone in acetabular reconstruction; a 90% survival was achieved at 10-15 years (10). The Exeter group (20) have been unable to match these results in a study following 111 acetabular reconstructions in which grafting was utilized. Excellent to good outcomes were recorded in 87% of patients at a mean follow-up of 6 years; 25% of cases showed some degree of acetabular migration. In order to overcome this comparatively high complication rate, more vigorous impaction in combination with a stronger mesh for bone containment was recommended; the utilization of bone chips > 1 cm was also recommended although not elaborated upon (14,20). The utilization of cementless cups in combination with impaction grafting in acetabular reconstruction is not widespread. This is centred on the belief that impacted allograft bone has no or limited bone ingrowth potential (4-6). Current dogma further suggests that at least 30-50% contact with host bone is required and that the stiff, cementless cups with incite bone resorption (stress shielding) through the disparity in elastic moduli (17). Trabecular metal _ has a porosity similar to bone and provides an environment more favorable to bone graft remodeling than conventional metals (7,16,17); it has also been shown to have excellent ingrowth potential (19,21). The current pilot study has sought to evaluate the efficacy of TM in combination with impaction grafting of autograft bone in the repair of a simulated Type III acetabular defect using a sheep model. Materials And MethodsThe experimental protocol was submitted to, and approved by, the institutional Animal Welfare Committee. A unilateral total hip replacement (THR) was performed using a 22 mm (inner diameter) polyethylene cup and a cemented, modular femoral stem and head using an already established protocol (22). With the femoral head removed a 2 cm diameter defect was created in the cranio-dorsal aspect of the left acetabulum. A retainment mesh was fashioned and secured to the labrum cranial to the acetabulum with 3.5 mm screws. Milling of the excised femoral head provided cortico-cancellous bone (4 mm chip size) which was then impacted into the defect created. Subsequently a 1.5 cm TM plug was placed onto the impacted bone surface; this plug had no contact with the surrounding host bone (Figure 1).
First generation cementing techniques were then applied to both the polyethylene cup and the femoral stem. The sheep was recovered in a sling for the first day post-operatively after which normal weight bearing ensued. Analgesia was provided for three days postoperatively. At one week post-operatively the sheep was returned to pasture for the duration of the study. Evaluative methods employed included radiography, macrophotographs, microradiology and histology. Radiographs were obtained post-operatively and at 6 and 8 weeks. At 7 weeks the sheep dislocated the reconstruction; this was reduced successfully under general anaesthesia. The dislocation recurred at 8 weeks and the sheep was euthanatized. (Figure 2)
Five mm transverse sections were taken spanning the width of the reconstruction and placed in 10% Buffered Formalum solution. Each section was photographed for a visual assessment of bone ingrowth. Preserved sections underwent micro-radiography allowing assessment of bone-implant interface occurrences. Histological assessment included the administration of a fluorochrome (calcein green) at 6 weeks postoperatively to enable visualization of new bone formation using fluorescent microscopy. Unstained ground sections were assessed using fluorescence microscopy using a Leica DM 6000 B fluorescence microscope attached to the Quantimet Version 550 IW image analysis system. Following fluorochrome analyses, ground sections were stained with toluidine blue and examined using plain light microscopy on an Olympus BH-2 microscope (Olympus Optical Company, Japan). Representative digital microphotographs of the histology were taken. ResultsNo untoward effects of the procedure were observed post-operatively with the sheep released to graze at pasture at one week. The first (calcein green) of three fluorochrome injections was performed at 6 weeks uneventfully. At 7 weeks the animal was observed to be lame. Radiographic evaluation under general anaesthesia revealed a dislocation of the femoral components with the femoral head resting cranial and dorsal to the acetabulum. This was reduced manually under traction and the sheep recovered. Normal weight bearing followed. At 8 weeks post-operatively, lameness was again observed. Radiographs (Figure 2) revealed a recurrence of the dislocation; the sheep was euthanatized and appropriate specimens retrieved. Macrophotography: Digital macrophotographs of salient sections were recorded and enable some assessment of the extent of bone ingrowth and/or fibrous tissue formation along the implant-autograft interface. Sections removed at the periphery of the TM implant consistently displayed a degree of fibrous tissue invasion at the interface between implant and autograft bone (Figure 3).
Through the majority of the sections however, little or no fibrous tissue was observed at the interface (Figure 4).
Figure 5 is a higher magnification view showing cement penetration into the TM and an implant-autograft interface with no fibrous tissue evident.
Microradiography: Microradiographs obtained concur with the findings with macrophotographs. Through the majority of sections little or no fibrous tissue formation was observed at the implant-autograft interface (Figure 6).
Histology: Assessment of sections under fluorescent microscopy highlighted significant amounts of new bone formation within the autograft bed, at the implant-bone interface and within the implant itself (Figure 7,8).
Assessment of the same sections under light microscopy and stained with Toluidene blue confirmed the presence of new bone formation at all three locations (autograft, interface, implant). Bone ingrowth into the outer porous spaces (1-1.5 mm depth) of the TM implant was observed along most of the surface to varying degrees (Figures 9).
DiscussionThe existing bone stock and type of defect are significant factors in surgical decision making when undertaking revision of acetabular components. The intent is to restore anatomy and to achieve stable fixation of the new acetabular component (s). Unlike more traditional metal options, trabecular metal has a porosity akin to bone and confers a more favourable mechanical environment in which bone graft materials (depending on their constitution) can function; this may be osteoconductive, osteoinductive or combination thereof (7,16,17,19,21). The current pilot study has provided an opportunity to evaluate the utilization of a Trabecular Metal implant when used in conjunction with impaction of autologous bone graft material and cement fixation of components; to this end the model has proven satisfactory. The literature (5,9) highlights concerns regarding the osseointegration of porous metal components when pressfit into a bone-deplete acetabulum. Recent studies (7,8) have evaluated the incorporation of primarily allogenic bone graft material to fill the defect followed by the placement of a cemented polyethylene cup. The current study has sought to evaluate the utilization of TM to augment the repair when used in combination with bone graft material and cemented components. Strong evidence of the early ingrowth of new bone into a trabecular metal implant has been observed, in this study, as early as six weeks post-operatively. Microradiography and histological techniques have shown good continuity of direct bone graft apposition and ingrowth along the majority of the trabecular metal implant. New bone ingrowth into the TM was observed extending as much as 1-2 mm . This is observed in the amount of fluorescent label apparent within the TM (Figure 9) and is corroborated by lightmicroscopy (Figure 7,8). This augers well for the provision of a stable mechanical environment in which continued ingrowth and increasing strength of the fixation is anticipated. The current study was intended to proceed over a three month period with fluorochromes administered at 6, 8 and 10 weeks. Repeat dislocation necessitated euthanasia of the animal prematurely; it was believed the dislocation was a response to a significantly modified polyethylene cup which was quite shallow. Based on the significant presence of fluorochrome label at six weeks we believe all three labels, had they been administered, would have been present in the trabecular metal indicating ongoing new bone ingrowth and complimentary strength of the fixation. Anticipated future studies will involve the use of an uncemented TM cluster cup in which migration of the acetabular components will be measured using radiostereometric analysis in combination with microradiography and histological assessments. This preliminary study has demonstrated that trabecular metal can be successfully used in impaction grafting in the absence of direct contact with peri-prosthetic host bone. New bone deposition was observed at depths of 1-3 mm inside the TM interface, in this study, as early as 6-8 weeks. This augers well for the utilization of trabecular metal to augment acetabular reconstruction in Type III scenarios. We now anticipate, that over longer periods of time, with continuing ingrowth and increasing strength, the provision of a stable mechanical environment will be achieved in response to the surgical techniques employed in this model. AcknowledgementsThe authors would like to acknowledge the contributions of Margaret McGee, Department of Orthopaedics and Trauma, Royal Adelaide Hospital for histological preparation and Peter Self, Adelaide Microscopy, for microradiography and micro CT. References1. D’Antonio JA, Capello WN, Borden LS, Bargar WL, Bierbaum BF, Boettcher WG, Steinberg ME, Stulberg SD, Wedge JH: Classification and management of acetabular abnormalities in total hip arthroplasty. Clin Orthop Relat Res 1989, 243: 126-137. (s) 2. D’Antonio JA: Periprosthetic bone loss of the acetabulum. Classification and management. Orthop Clin North Am 1992, 23(2): 279-290. (s) 3. D’Antonio JA: Acetabular reconstruction in revision total hip arthroplasty. Semin Arthroplasty 1995, 6(2): 45-59. (s) 4. Brubaker SM, Brown TE, Manaswi A, Mihalko WM, Cui O, Saleh KJ: Treatment options and allograft use in revision total hip arthroplasty the acetabulum. J Arthroplasty 2007, 22(7 Suppl 3): 52-56. (s) 5. Jasty M: Jumbo cups and morselized graft. Orthop Clin North Am, 1998, 29(2): 249-254. (s) 6. Whaley AL, Berry DJ, Harmsen WS: Extra-large uncemented hemispherical acetabular components for revision total hip arthroplasty. J Bone Joint Surg Am 2001, 83-A(9): 1352-1357. (s) 7. Gross AE, Goodman S: The current role of structural allografts and cages in revision arthroplasty of the hip. Clin Orthop Relat Res 2004, 429: 193- 200. (s) 8. Comba F, Buttaro M, Pusso R, Piccaluga F: Acetabular reconstruction with impacted bone allografts and cemented acetabular components: a 2- to 13- year follow-up study of 142 aseptic revisions. J Bone Joint Surg Br 2006, 88(7): 865-869. (s) 9. Hendrich C, Engeimaier F, Mehling I, Sauer U, Kirschner S, Martell JM: Cementless acetabular reconstruction and structural bone-grafting in dysplastic hips. Surgical technique. J Bone Joint Surg Am 2007, 89 Suppl 2: 54-67. (s) 10. Slooff TJ, Huiskes R, van Horn J, Lemmens AJ: Bone grafting in total hip replacement for acetabular protrusion. Acta Orthop Scand 1984, 55(6): 593-596. (s) 11. Klein GR, Parvizi J, Shastri S, Frank HG, Sharkey PF, Hozack WJ: Simultaneous primary total hip arthroplasty and contralateral revision hip arthroplasty: role for the use of femoral head autograft. J Arthroplasty 2006, 21(4): 527-532. (s) 12. Schreurs BW, Bolder SB, Gardeniers JW, Verdonschot N, Sliiff TJ, Veth RP: Acetabular revision with impacted morselized cancellous bone grafting and a cemented cup. A 15- to 20-year follow-up. J Bone Joint Surg Br 2004, 86(4): 492-497. (s) 13. Schreurs BW, Busch VJ, Welten ML, Verdonschot N, Sloof TJ, Gardeniers JW: Acetabular reconstruction with impaction bone-grafting and a cemented cup in patients younger than fifty years old. J Bone Joint Surg Am 2004, 86(11): 2385-2392. (s) 14. Schreurs BW, Slooff TJ, Gardeniers JW, Buma P: Acetabular reconstruction with bone impaction grafting and a cemented cup. Clin Orthop Relat Res 2001, 393: 202-215. (s) 15. Somford MP, Bolder SB, Gardeniers JW, Slooff TJ, Schreurs BW. Favorable survival of acetabular reconstruction with bone impaction grafting in dysplastic hips. Clin Orthop Relat Res 2008, 466(2): 359-365. (s) 16. Gross AE, Goodman SB: Rebuilding the skeleton: the intraoperative use of trabecular metal in revision total hip arthroplasty. J Arthroplasty 2005, 20(4 Suppl 2): 91-93. (s) 17. Gross AE: Restoration of acetabular bone loss 2005. J Arthroplasty 2006, 21(4 Suppl 1): 117-120. (s) 18. Bolder SB, Melenhorst J, Gardeniers JW, Slooff TJ, Veth RP, Schreurs BW: Cemented total hip arthroplasty with impacted morselized bone-grafts to restore acetabular bone defects in congenital hip dysplasia. J Arthroplasty 2001, 16(8 Suppl 1): 164-169. (s) 19. Boscainos PJ, Kellett CF, Maury AC, Backstein D, Gross AE: Management of periacetabular bone loss in revision hip arthroplasty. Clin Orthop Relat Res 2007, 465: 159-165. (s) 20. Timperley J, Bannister G, Gie G, Ring P, Ling R: Effect of cup geometry and the presence of cement on acetabular component fixation. Arch Orthop Trauma Surg 1992, 111(6): 301-304. (s) 21. Nehme A, Lewallen DG, Hanssen AD: Modular porous metal augments for treatment of severe acetabular bone loss during revision hip arthoplasty. Clin Orthop Relat Res 2004, 429: 201-208. (s) 22. Field JR, Aberman H, Carbone A, Sharpe P, Howie DW: An Ovine model for Total Hip Replacement: Operative procedure and complications. Vet Comp Orthop Traumatol, 2001, 14: 32-39 (s) This article was last modified on Thu, 24 Sep 09 13:13:14 -0500 This page was generated on Sun, 21 Mar 10 02:56:45 -0500, and may be cached. |
|