The Internet Journal of Anesthesiology™ ISSN: 1092-406X| Home | Editors | Current Issue | Archives | Instructions for Authors | Disclaimer |Combined use of stylet and introducer tool in Proseal™ laryngeal mask aids insertion in halo frame immobilized patient
Rashid M. Khan M.D.
Pradeep K. Sharma M.D.
Naresh Kaul M.D.
Ashok Sumant M.D., FFA
Citation: R. M. Khan, P. K. Sharma, N. Kaul & A. Sumant : Combined use of stylet and introducer tool in Proseal™ laryngeal mask aids insertion in halo frame immobilized patient . The Internet Journal of Anesthesiology. 2008 Volume 18 Number 2 Keywords: Proseal™ laryngeal mask airway | halo frame | cervical spine immobilization Table of ContentsAbstractSuccessful insertion of Proseal™ laryngeal mask airway requires placement of patient's head and neck in a 'sniffing' position which is not attainable in patients with halo frame. We attempted to see if shaping the Proseal™ laryngeal mask airway into a 'C” aids its insertion in a patient with halo frame. IntroductionSuccessful insertion of Proseal ™ laryngeal mask airway (PLMA) requires placement of patient's head and neck in a ‘sniffing' position 1 . Unfortunately this position is not attainable in patients with halo frame for immobilization of cervical spine. We recently had a patient with multiple trauma needing open reduction of the # right tibia. Patient had also sustained # C2 which had been immobilized with a halo frame. Insertion of PLMA failed despite two attempts by an experienced anaesthetist with and without the aid of an introducer tool. However, PLMA placement turned out to be easy by contouring it into a C-shape configuration with the help of stylet placed in the drain tube. Case reportA 55 year-old, ASA grade II male patient [approximate weight 70 kg] with multiple fractures, was posted for elective open reduction of # right tibia. The patient had been put on a halo frame for # C2. The patient requested for all procedures to be done under general anaesthesia. Except for a mild chronic obstructive pulmonary disease [COPD], the patient had no other medical disease. All investigations were within acceptable limits. The patient was premedicated with inj pethidine 50 mg and inj ondansetron 4 mg administered intramuscularly 30 min before induction of anaesthesia. Following 3 min preoxygenation, anaesthesia was induced with propofol 120 mg and fentanyl 75 µg intravenously. After ascertaining easy mask ventilation, neuromuscular blockade was achieved with suxamethonium 50 mg administered intravenously. Keeping in view the difficult tracheal intubation and the possible need for a higher ventilatory pressure in a patient of COPD, PLMA was selected as the airway device for this patient. After adequate relaxation, #4 PLMA was introduced but failed to be correctly placed. A second attempt with the introducer tool resulted in PLMA placement with some difficulty. A high peak inflation pressure of 28 cmH2O was noted with a tidal volume of 550ml. High inflation pressure and failure to pass a nasogastric tube via the drain tube suggested folding back of the PLMA tip. It was reasoned that folding back of the PLMA probably occurred as the PLMA tip met the posterior pharyngeal wall at 90° or <90°. To overcome this problem, a well lubricated large Portex ™ stylet was introduced via the drain tube till its distal end and a C-shaped configuration of the PLMA was made [Figure 1].
Withdrawal of the stylet by 4-5 cm restored the PLMA shape to original [Figure 2].
With the stylet and introducer in place, the reconfigured PLMA was easily introduced (Figure 3).
As soon as the device negotiated the oropharyngeal curve, and before going any further, the stylet was withdrawn by about 4-5 cm (Figure 4) and the device was smoothly placed.
Both the stylet and the introducer tool were now removed (Figure 5) and cuff inflated with 20 ml air.
Good ventilation was recorded as evidenced by chest auscultation, capnographic tracing and a low inflation pressure of 21 cmH2O. Adequate patency of drain tube was ascertained by the easy passage of a nasogastric tube. Throughout the procedure heart rate, mean arterial pressure and oxygen saturation ranged between 74 – 81/min, 88 – 97 mmHg, and 97-98% respectively. Time to insert the PLMA with the aid of introducer and the stylet was <10 seconds. Patient's consent to publish his photograph has been obtained and uploaded along with this article. DiscussionIn addition to the conventional digital and introducer tool techniques, several other methods have been used to facilitate insertion of the PLMA. These include railroading PLMA over gum elastic bougie 2,3 or the flexible fiberscope 4 , and gastric tube 5 or suction catheter 6 guided insertion of the PLMA. In the present case, after the initial failure of PLMA placement with digital and introducer tool techniques, the above mentioned techniques were considered but not applied. The disadvantage of the gum elastic bougie is that it is rigid and requires laryngoscope guidance which would have been difficult in this case. Suction catheter and gastric tube are soft and may have failed to guide the PLMA around the oropharynx held fixed in the halo frame. Furthermore, successful use of the above techniques has not been reported in patients with halo frame immobilization of the head and neck. Using our technique, we could curve the distal portion of the PLMA. By passing a well lubricated stylet into the drain tube, we could shape the PLMA into a gentle C-shape and yet restore its shape by withdrawing the stylet by 4-5 cm (Figures 1 and 2). The distal curvature extending up to the mask tip made the negotiation of the oropharyngeal curvature easier in an otherwise fixed head and neck position. Furthermore, keeping the distal end of the stylet till the tip of the drain tube prevented the folding back of the PLMA as it met the posterior pharyngeal wall. Withdrawing the stylet by approximately 4-5 cm after negotiating the oropharyngeal curve, restored the original shape of the device without impeding progress of the tip of the PLMA towards the upper oesophageal opening as confirmed by smooth passage of the nasogastric tube subsequently. In conclusion, a premounted PLMA with introducer tool and lubricated stylet facilitates its easy placement in patients whose head and neck has been immobilized in a halo frame. Address for CorrespondenceDr. Naresh Kaul References1. Cook TM, Lee G, Nolan JP. The ProsealTm laryngeal mask airway: a review of the literature. Can J Anesth 2005; 52: 739-760. (s) 2. Howarth A, Brimacombe J, Keller C. Gum elastic bougie-guided insertion of the Proseal laryngeal mask airway. A new technique. Anaesth Intens Care 2002; 30: 624-627. (s) 3. Brimacombe J, Keller C, Vosoba Judd D. Gum elastic bougie-guided insertion of the ProsealTm laryngeal mask airway is superior to the digital and introducer tool techniques. Anesthesiology 2004; 100: 25-29. (s) 4. Brimacombe J, Keller C. Awake fiberoptic-guided insertion of the ProSeal laryngeal mask airway. Anaesthesia 2002; 57: 719. (s) 5. Drolet P, Girard M. An aid to correct positioning of the Proseal laryngeal mask. Can J Anesth 2001; 48: 718-719. (s) 6. Gracia-Aguado R, Vinoles J, Brimacombe J, Vivo M, Lopez-Estudillo R, Ayala G. Suction catheter guided insertion of the ProSealTm laryngeal mask airway is superior to the digital technique. Can J Anesth 2006; 53: 398-403. (s) This article was last modified on Fri, 13 Feb 09 13:16:30 -0600 This page was generated on Tue, 09 Feb 10 11:34:08 -0600, and may be cached. |
|